Graphtec GL840

Testing Recording and Analize

Three types of input system enable to measure various phenomena
Input system 1 : Multifunction analog input ports
Contains a highly isolated input system which ensures that signals are not corrupted by noise from other channels. The GL840`s inputs are suitable for combined measurements from voltage,temperature,humidity, logic, and pulse signals.
The standard configuration has 20 analog input channels. It is expandable to 200 channels by adding optional 20 channels extension terminal base unit(B-566) and input terminal units(B-564 or B-565).

| Withstand voltage & Accuracy | Multi-input type (B-564) | Withstand-voltage type (B-565) | |
|---|---|---|---|
| Voltage | Input voltage range | 20 mV to 100V | 20mV to 100V |
| Max. voltage (Input - GND) | 60 Vp-p | 300 Vp-p | |
| Temp. | Thermocouple | R, S, B, K, E, T, J, N, W (WRe5-26) | |
| RTD (Resistance Temp. Detector) | Pt100 (IEC751), JPt100 (JIS), Pt1000 (IEC751) | ||
| Accuracy | Voltage | ± 0.1% of F.S. | ±(0.05% of FS + 10μV) |
| Temperature* | ± 1.55 ºC | ± 1.1 ºC | |
* Accuracy rating for K-type thermocouple at 100˚C includes reference junction compensation. Accuracy varies by the temperature levels and thermocouple types.
The following shows how a standard configuration is expanded to more than 40 channels

Configuration for additional channels
|
Number of channels |
20 channels |
40 channels |
100 channels |
200 channels |
|---|---|---|---|---|
|
GL840 unit (GL840-M or GL840-WV) |
1 set |
1 set |
1 set |
1 set |
|
Connection cable(B-567-05 or -20) |
N/A |
1 pc |
1 pc |
1 pc |
|
Terminal base (B-566) |
N/A |
2 sets |
5 sets |
10 sets |
|
Input terminal (B-564 or B-565 ) |
N/A |
1 set |
4 sets |
9 sets |
Offers longer cable for the input terminals
Input terminal blocks can be connected directly(in daisy chain), or using the B-565 cable(s). This allows the input terminals to be placed in separate locations according to the need of the application.
The input terminal unit and the GL840 main body can be extended by using an extended connections cable.
*If the signal is affected by noise, It may be required to use a slower sampling.

Input system 2 : Support dogotal sensors
Digital sensors and input terminal/adapters for the GL100 connects to the GL840 directly

Input system 3 : 4 channels of Logic/Pulse inputs
Supports 4-channels logic or pulse signal inputs. Pulse mode allows cumulative, instant, or rotational values for industrial measurement capability with speed and flow.
Maximum sampling interval of up to 10ms
Provides faster sampling rates for voltage measurements. You are able to achieve up to 10ms sampling speed when limiting the number of channels in use.
|
Sampling Interval |
10ms | 20ms | 50ms | 100ms | 200ms | 500ms | 1s | 2s | |
|---|---|---|---|---|---|---|---|---|---|
|
Number of channel |
1 | 2 | 5 | 10 | 20 | 50 | 100 | 200 | |
|
Measuring |
Voltage |
● | ● | ● | ● | ● | ● | ● | ● |
|
Temperature |
- | - | - | ● | ● | ● | ● | ● | |
Supports large-size SD memory card for reliable longterm measurement
New GL840 series carries two SD memory card slots for storage device. The SDHC type SD memory card is supported up to 32GB. 4GB SD memory card comes as a standard accessory installed in the first slot.
the captured data can be stored in GBD(Graphtec binary data) or CSV file format.
Capturing time* (when all 20 analog channels are being used with Logic/`Pulse inputs turned off.)
|
Sampling |
10ms | 50ms | 100ms | 200ms | 500ms | 1s | 10s |
|---|---|---|---|---|---|---|---|
|
GBD format |
31 days | 77 days | 95 days | 108 days | 270 days | Over 365 | Over 365 |
|
CSV format |
3 days | 11 days | 16 days | 21 days | 54 days | 109 days | Over 365 |
*Figures are approximate. File size of captured data is 2GB in GBD or CSV file format on this chart.
Sampling interval is limited by the number of channels in use.(10ms:1ch, 50ms:5ch, 100ms:10ch)
Limited sampling speed is available when digital sensors and GL100-WL are used as a remote monitoring device.
What you can do using the wireless LAN (option)
Case 1 : Combining GL100-WL and GL840
GL100-WL can now be connected to the GL840 as a remote sensor using the WLAN feature.
You can expand your measurement variety by adding the sensors available on the GL100-WL unit.
The measured value by GL100-WL will appear in a single file along with the measurement value from the GL840 main inputs. GL840 will now take in direct information from the GL100-WL units.GL840 supports up to five units of the GL100-WL.

Application Example : Air conditioning performance test
Easily get the additional information that is required for the multi channel temperature measurement.

Measurement data from GL100-WL, GS sensor on the main unit can be displayed as a waveform and digital value.
Case 2 : Remote monitoring of GL840 by PC or Smart device

Measured data can be monitored and controlled via PC or a smart device using the applicaiton software.
Using a wireless LAN, there is no needs to connect by USB cable with the GL840.

Available configuration

Available functions
| Display the data | Control the Start / Stop | Change the settings | Saves the data | Receives warning mail | |
|---|---|---|---|---|---|
| Smart Device |
YES | YES | YES | N/A | N/A |

Available functions
| Display the data | Control the Start / Stop | Change the settings | Saves the data | Receives warning mail | |
|---|---|---|---|---|---|
| PC | YES | YES | YES | YES | N/A |

Available functions
| Display the data | Control the Start / Stop | Change the settings | Saves the data | Receives warning mail | |
|---|---|---|---|---|---|
| Smart Device |
YES | YES | YES | N/A | N/A |

Available functions
| Display the data | Control the Start / Stop | Change the settings | Saves the data | Receives warning mail | |
|---|---|---|---|---|---|
| Smart Device |
YES | YES | YES | N/A | N/A |
| PC | YES | YES | YES | YES | N/A |
Case 3 : Remote monitoring of GL840 using Network

GL840 can be connected to a LAN (Local Area Network) via an wireless LAN access point.
Available configuration

Available functions
| Display the data | Control the Start / Stop | Change the settings | Saves the data | Receives warning mail | |
|---|---|---|---|---|---|
| Smart Device |
YES | YES | YES | N/A | N/A |
| PC | YES | YES | YES | YES | YES |

Available functions
| Display the data | Control the Start / Stop | Change the settings | Saves the data | Receives warning mail | |
|---|---|---|---|---|---|
| Smart Device |
YES | YES | YES | N/A | YES |

Available functions
| Display the data | Control the Start / Stop | Change the settings | Saves the data | Receives warning mail | |
|---|---|---|---|---|---|
| PC | YES | YES | YES | YES | YES |
Case 4 : Remote monitoring of GL840 via Internet

Using a router that supports a dinamic DNS service,
it enables to monitor GL840 remotely via Internet.
In other words, it is possible to contact your GL840 whereever you are !!
Available configuration

■Available functions
| Display the data | Control the Start / Stop | Change the settings | Saves the data | Receives warning mail | |
|---|---|---|---|---|---|
| Smart Device |
N/A | N/A | N/A | N/A | YES |
| PC | N/A | N/A | N/A | N/A | YES |

Available functions
| Display the data | Control the Start / Stop | Change the settings | Saves the data | Receives warning mail | |
|---|---|---|---|---|---|
| Smart Device |
YES | YES | YES | N/A | YES |
| PC | YES | YES | YES | YES | YES |
Ring capture function
The most recent data is saved when the memory is configured in ring memory mode.(Number of capturing data is 1000 to 2000000 points)
Relay capture function
Data is continuously saved to multiple files up to 2GB without losing any data until capturing is stopped when the memory is configured in the relay mode.
Hot-swapping the SD memory card
SD card can be replaced during data capturing when the sampling interval is 100ms or slower.
Back up recording data
The data file can be created periodically for purpose of backing up data while recording it, and the file is automatically transferred to another storage device on the unit or FTP server on the network.
Backup interval: 1, 2, 6, 12, 24 hour(s)Backup file destination: SD memory card or FTP server
Data file format: GBD (binary) or CSV (text)
• Available sampling speed is the 100 ms or slower when using the CSV format.
• When the RING mode or external pulse synchronization sampling is selected for recording, the backup function is not available.
• The storage device specified as the recording destination of the measurement data can not be set as the transfer destination of the backup file.
• When backup is enabled and data file format is specified with CSV format, SD memory card exchange (hot-swapping) and RELAY recording are not available.
Data file verification
It is a function to check the integrity of the data file when replaying the data file. It compares check-sum generated from the data file and recorded check-sum. When two check-sums do not match, the data file may be corrupted or has been changed by someone. The check-sum code is recorded into the file when data file is created by the GL unit.
* Check-sum code is added to normal data recording file and also file created for backup, ring or relay recording. Adding check-sum is available for data files created in GBD format.When the recorded data file is converted and saved using GL100_240_840-APS or GL-Connection software, the check-sum code will not match.
* This function is available with firmware version 1.43 or rater.
Useful functions
Displays the data by a bar chart (in firmware ver.1.10 or later)
The integrated data that is measured by the digital sensors can be displayed by a bar chart in the GL840 series.
| Available GS sensors | Measurement | Unit |
| GS-TH (Temp/Humidity) sensor | Integrated temperature | ℃h |
| GS-DPA-AC adapter with GS-Acxxx sensor (AC current/power) | Integrated power | kWh |
| GS-LXUV (Illuminance/UV) sensor | Integrated illuminance | lxh or klxh |
| Integrated UV | mW/cm2h |
Alarm output function
Based on set conditions for each channels, alarm signal can be placed using the four channel alrm out put ports.*
*Input/Output cable(B-513 option) is required to connect the alrm output ports to external buzzer/light mechanism.
USB drive mode
USB drive mode function enables data to be transferred to the PC from GL840 by drag&drop feature.
Navigation function
Simple to use navigation screen allows setting operatin for measurement and wireless LAN adapter.
Large easy-to-read 7-inch wide color LCD
Carries a clear 7-inch wide TFT color LCD screen (WVGA:800x480 dots) for the GL840. Monitoring data are displayed in waveform or digital form option. Parameter settings can be displayed on the screen.
3 Types of power source
Choose from AC power supply, DC supply* or the rechargeable battery pack.*
*DC power drive cable(B-514) and battery pack(B-569) are optional accessories.
Networking features
WEB&FTP server function
GL840 can be controlled externally via a network on the WEB browser, which also supports monitoring and transfer of signals and captured data.
FTP client function
Captured data is periodically transferred to the FTP server for backup.
NTP client function
The clock on the GL840 is periodically synchronized with the NTP server.
*The GL840 needs tobe connected to a LAN enviroment using the available Ethenet/WLAN ports for above functions.
Cover (B-578)


Protect the main body from cosmetic damages and minor impacts.
* The cover is not to protect from hard impact.
Storage case (B-581)
Portable case to store GL840 and signal input cables for easy handling.
* The case is not to protect from hard impact.
GL840 Main unit specifications
| Item | Description | ||
|---|---|---|---|
| Model number | GL840-M | GL840-WV | |
| Number of analog input channels | 20 channels in standard configuration, Expandable up to 200 channels | ||
| Number of analog input terminals | Up to 10 terminals (20 channels / terminal), standard config: 1 | ||
| Type of analog input terminal | Multi-input type, Withstand-voltage type | ||
| Port for digital sensor | 1 port for the sensor/input terminal/adapter of the GL100 | ||
| External input/ output *1 |
Input *2 | Trigger or Sampling (1 channel), Logic/Pulse (4 channels) | |
| Output *3 | Alarm (4 channels) | ||
| Sampling interval | 10 ms to 1 hour (10ms to 50ms: voltage only) *4, External signal | ||
| Time scale of waveform display | 1 sec. to 24 hour /division | ||
| Trigger, Alarm function |
Trigger action | Start or stop capturing data by the trigger | |
| Repeat action | Off, On (auto rearmed) | ||
| Trigger source | Start: Off, Measured signal, Alarm, External, Clock, Week or Time Stop: Off, Measured signal, Alarm, External, Clock, Week or Time |
||
| Condition Setting | Combination: OR or AND Analog signal: Rising (High), Falling (Low), Window-in, Window-out Logic signal: Pattern (combination of each input signal in high or low) Pulse (number of count): Rising (High), Falling (Low), Window-in, Window-out |
||
| Alarm output | Outputs a signal when alarm condition occurs in the input signal *5 | ||
| Pulse input function |
Rotation count (RPM) mode |
Counts the number of pulses per sampling interval and converts to rpm (rotations per minute), Number of pulses for one rotation can be set to 50, 500, 5000, 50k, 500k, 5M, 50M, 500M rpm/F.S. (rpm./Full Scale) |
|
| Accumulating count mode |
Accumulates the number of pulses from the start of measurement 50, 500, 5000, 50k, 500k, 5M, 50M, 500M C/F.S. (Counts/Full Scale) |
||
| Instant count mode |
Counts the number of pulses per sampling interval 50, 500, 5000, 50k, 500k, 5M, 50M, 500M C/F.S. (Counts/Full Scale) |
||
| Calculation function |
Between channels | Addition, Subtraction, Multiplication, and Division for analog input | |
| Statistical | Select two calculations from Average, Peak, Maximum, Minimum, RMS | ||
| Search function | Search for analog signal levels, values of logic or pulse or alarm point in captured data |
||
| Interface to PC | Ethernet (10 BASE-T/100 BASE-TX), USB (Hi-speed), WLAN (using B-568 option) | ||
| Storage device |
Media | SD memory card (Support SDHC, up to 32 GB), supports 2 slots *6 | |
| Saved contents | Captured data, Setting conditions, Screen copy | ||
| Capturing mode | Mode: Normal, Ring, Relay Ring: Saves most recent data (Number of capturing data: 1000 to 2000000 points) *7 Relay: Saves data to multiple files without losing data until dada capturing is stopped |
||
| Replay data | Replays captured data that was saved in the GL840 (in GBD or CSV format) | ||
| Scaling (Engineering unit) function | Measured value can be converted to specified engineering unit • Analog voltage: Converts using four reference points (gain, offset) • Temperature: Converts using two reference points (offset) • Pulse count: Converts using two reference points (gain) |
||
| Action during data capture | • Displaying past data (using dual display mode (Current + Past data)) • Hot-swapping the SD memory card • Saving data in between cursors |
||
| Display | Size | 7-inch TFT color LCD (WVGA: 800 x 480 dots) | |
| Language | English, French, German, Chinese, Korean, Russian, Spanish, Japanese | ||
| Information *8 | Waveform in Y-T with digital values, Waveform only, Digital value, Digital values and statistics values |
||
| Operating environment | 0 to 45 ºC, 5 to 85 % RH (non condensed) (When operating with battery pack 0 to 40 ºC, charging battery 15 to 35 ºC) |
||
| Power source | AC adapter | 100 to 240 V AC, 50/60 Hz (1 pc of adapter is attached as standard accessory) | |
| DC power | 8.5 to 24 V DC (DC drive cable (option B-514) is required) | ||
| Battery pack | Mountable two battery packs (battery pack (option B-517): 7.2V DC, 2900mAh) | ||
| Power consumption *9 | Max. 38 VA | ||
| External dimensions (W x D x H in mm, Excluding projections |
Approx. 240 x 158 x 52.5 | Approx. 240 x 166 x 52.5 | |
| Weight *10 | Approx. 1010 g | Approx. 1035 g | |
*2 Input signal;
- Voltage range: Up to 24V (common ground)
- Signal type: Voltage, Open collector, Contact (relay)
- Threshold: Approx. + 2.5 V (Hysteresis: Approx. 0.5V (2.5V to 3V))
- Voltage: Max. 30V,
- Current: Max. 0.5A,
- Collector dissipation: Max. 0.2W
*5 Output port can be specified in each input channel.
*6 4GB SD memory card is installed to slot 1 as standard accessory.
*7 Size of the capture data will be limited to 1/3 of available memory.
*8 Display mode is switched every time the dedicated key is pressed. In magnified digital value mode, the
displayed channel number can be specified. In the waveform display mode, the changing of the time scale will be effective from the point of the next displayed data.
*9 Rating under maximum power consumption using the AC adapter, with LCD display on, and battery pack(s)
being charged.
*10 Excludes AC adapter and battery pack.
Software specifications for PC
| Item | Description | ||
|---|---|---|---|
| Supported OS | Windows 8.1, 8, 7, Vista (32/64-bit edition) | ||
| Supported device | GL840 (USB, Ethernet, WLAN), GL100 (USB, WLAN) | ||
| Functions | Control the GL series, Real-time data capture, Replay data, and Data format conversion | ||
| Supported units & channels | Up to 1000 channels total, Up to 4 groups (number of units is limited by model) | ||
| Settings control | Input condition, Capturing condition, Trigger/Alarm condition, Report, etc. | ||
| Capturing data | Saved to PC | Saves captured data in real time (in GBD binary or CSV format) | |
| Saved to GL unit | Saves to the SD memory card (in GBD binary or CSV format) | ||
| Displayed information | Y-T waveform, Digital values, Report, X-Y graph (specified period of data, data reply only), Two displays for the current and past data, and Statistical calculation |
||
| File operation | Converting data format to CSV from GBD binary, merge multiple data files in the time axis or as an additional channel |
||
| Warning function | Send e-mail to the specified address when the alarms occur | ||
| Statistical calculation | Maximum, Minimum, and Avarage during data capturing | ||
| Report function | Creates the daily or monthly report automatically | ||
Software specifications for Smart device
| Item | Description | |
|---|---|---|
| Model name | GL-Connect | |
| Supported OS | Android 4.1 to 4.4, iOS 7/8 | |
| Supported device | GL840 (WLAN), GL100 (WLAN) | |
| Functions | Control the GL series, Display measured data in waveform or digital value | |
| Supported units | Up to 10 units | |
| Settings control | Start/Stop, Sampling interval | |
| Capturing data | Saves captured data in the GL main body (data cannot be saved in the smart device) | |
| Displayed information | Data captured in real time by digital value, Replay the data stored in the GL body by the waveform | |
Wireless LAN unit (option) specifications
| Item | Description | |
|---|---|---|
| Model number | B-568 | |
| Supported device | GL840 | |
| Communication method | Wireless communication (using radio waves in the 2.4GHz band) | |
| Supported WLAN system | IEEE802.11b/g/n WPS: Push button or PIN method Security protocols: WEP64, WEP128, WPA-PSK/WPA2-PSK, AKIP/AES Communication distance: Approx. 40m (depending on the conditions of radio communication) |
|
| Installed location | Attached to the SD CARD slot number 2 on the GL840 * When the wireless LAN unit is installed, the SD memory card cannot be used in slot number 2 |
|
| Function | Access Point mode: Communicate with the GL100-WL as a remote sensor (captured data in the GL100-WL is transferred to GL840) Station mode: Communicate with PC or Smart device (control GL840 and transfer the data from GL840) |
|
| Connected number of GL100-WL | GL840: Up to 5 units of the GL100-WL | |
GL840 Analog input specifications
| Item | Description | ||||
|---|---|---|---|---|---|
| Model number | GL840-M, Input terminal B-564 | GL840-WV, Input terminal B-565 | |||
| Input method | All channels isolated balanced input *11, Scans channels for sampling | ||||
| Type of input terminal | Screw terminal (M3 screw) | ||||
| Measurement range |
Voltage | 20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20, 50, 100 V, and 1-5V F.S. (Full Scale) | |||
| Thermocouple | Type: K, J, E, T, R, S, B, N, W (WRe5-26) Range: 100, 500, 2000 ºC *12 |
||||
| RTD (Resistance Temperature Detector) |
Type: Pt100, JPt100 (JIS), Pt1000 (IEC751) Range: 100, 500, 2000 ºC *12 |
||||
| Humidity | 0 to 100 % RH - using the humidity sensor (option B-530) | ||||
| Filter | Off, 2, 5, 10, 20, 40 (moving average in selected number) | ||||
| Measurement accuracy *13 | |||||
| Voltage | ± 0.1% of F.S. (Full Scale) | ± (0.05% of F.S. + 10μV) | |||
| Temperature (Thermocouple) *14 | |||||
| Type | Measurement range (TS: Temp Sense) |
Measurement accuracy | Measurement accuracy | ||
| R | 0 ≤ TS ≤ 100 ºC | ± 5.2 ºC | ± 4.5 ºC | ||
| 100 < TS ≤ 300 ºC | ± 3.0 ºC | ± 3.0 ºC | |||
| 300 < TS ≤ 1600 ºC | ± (0.05% of rdg. + 2.0 ºC) | ± 2.2 ºC | |||
| S | 0 ≤ TS ≤ 100 ºC | ± 5.2 ºC | ± 4.5 ºC | ||
| 100 < TS ≤ 300 ºC | ± 3.0 ºC | ± 3.0 ºC | |||
| 300 < TS ≤ 1760 ºC | ± (0.05% of rdg. + 2.0 ºC) | ± 2.2 ºC | |||
| B | 400 ≤ TS ≤ 600 ºC | ± 3.5 ºC | ± 3.5 ºC | ||
| 600 < TS ≤ 1820 ºC | ± (0.05% of rdg. + 2.0 ºC) | ± 2.5 ºC | |||
| K | -200 ≤ TS ≤ -100 ºC | ± (0.05% of rdg. + 2.0 ºC) | ± 1.5 ºC | ||
| -100 < TS ≤ 1370 ºC | ± (0.05% of rdg. + 1.0 ºC) | ± 0.8 ºC | |||
| E | -200 ≤ TS ≤ -100 ºC | ± (0.05% of rdg. + 2.0 ºC) | ± 1.0 ºC | ||
| -100 < TS ≤ 800 ºC | ± (0.05% of rdg. + 1.0 ºC) | ± 0.8 ºC | |||
| T | -200 ≤ TS ≤ -100 ºC | ± (0.1% of rdg. + 1.5 ºC) | ± 1.5 ºC | ||
| -100 < TS ≤ 400 ºC | ± (0.1% of rdg. + 0.5 ºC) | ± 0.6 ºC | |||
| J | -200 ≤ TS ≤ -100 ºC | ± 2.7 ºC | ± 1.0 ºC | ||
| -100 < TS ≤ 100 ºC | ± 1.7 ºC | ± 0.8 ºC | |||
| 100 < TS ≤ 1100 ºC | ± (0.05% of rdg. + 1.0 ºC) | ± 0.6 ºC | |||
| N | -200 ≤ TS < 0 ºC | ± (0.1% of rdg. + 2.0 ºC) | ± 2.2 ºC | ||
| 0 ≤ TS ≤ 1300 ºC | ± (0.1% of rdg. + 1.0 ºC) | ± 1.0 ºC | |||
| W | 0 ≤ TS ≤ 2000 ºC | ± (0.1% of rdg. + 1.5 ºC) | ± 1.8 ºC | ||
| R.J.C. | ± 0.5 ºC | ± 0.3 ºC | |||
| Temperature (RTD) *15 | |||||
| Type | Measurement range (TS: Temp Sense) |
Accuracy | Accuracy | ||
| Pt100 | -200 ≤ TS ≤ 100 ºC | ± 1.0 ºC | ± 0.6 ºC | ||
| 100 < TS ≤ 500 ºC | ± 0.8 ºC | ||||
| 500 < TS ≤ 850 ºC | ± 1.0 ºC | ||||
| JPt100 | -200 ≤ TS ≤ 100 ºC | ± 0.8 ºC | ± 0.6 ºC | ||
| 100 < TS ≤ 500 ºC | ± 0.8 ºC | ||||
| Pt1000 | -200 ≤ TS ≤ 100 ºC | ± 0.8 ºC | ± 0.6 ºC | ||
| 100 < TS ≤ 500 ºC | ± 0.8 ºC | ||||
| A/D converter | Sigma-Delta type, 16 bits (effective resolution: 1/40000 of the measuring full range) | ||||
| Maximum input voltage |
Between (+) / (-) terminal |
20 mV to 2 V range: 60 Vp-p, 5 V to 100 V range: 110 Vp-p |
|||
| Channels ((-) / (-)) | 60 Vp-p | 600 Vp-p | |||
| Channel / GND | 60 Vp-p | 300 Vp-p | |||
| Max. voltage (withstand) |
Between channels | 350 Vp-p (1 minute) | 600 Vp-p | ||
| Channel / GND | 350 Vp-p (1 minute) | 2300 Vrms AC (1 minute) | |||
*12 If the specifications of the temperature sensor is lesser or greater than the selected measurement range, GL840 can measure up to the specifications of the sensor.
*13 Subject to the following conditions:
- Room temperature is 23 oC ± 5 oC.
- When 30 minutes or more have elapsed after power has turned on.
- Filter is set to 10.
- Sampling rate is set to 1 sec, using 20-channel in GL840-M and 10-channel in GL840-WV.
- GND terminal is connected to ground.
*15 Supports 3-wire type sensor.
| Models | |
| GL840-M E | Data-logger GL840 cu 20 canale - precizie: ± 0.1%, suportă max. 110V pe intrare In stock |
| GL840-WV | Data-logger GL840 cu 20 canale - precizie: ± 0.05% + 10μV, suportă max. 300V pe intrare |
| Optional accessories | |
| B-564 | Terminal 20 canale intrare (necesită B-566) |
| B-565 | Terminal 20 canale intrare, rezistent la tensiune înaltă (necesită B-566) |
| B-566 | Bază pentru terminal intrare tip B-564 sau B-565 |
| B-568 | Adaptor WLAN, IEEE 802.11 b / g / n |
| B-567-05 | Cablu conexiune pentru baza extensie B-566, 50 cm - min. 1 buc necesară pentru a monta extensii canale |
| B-567-20 | Cablu conexiune pentru baza extensie B-566, 2m lungime - min. 1 buc necesară pentru a monta extensii canale |
| B-569 | Acumulator Litiu-Ion (7.2 V, 2900mAh) |
| B-530 | Senzor de umiditate - cablu 3m |
| B-551-10 | Rezistență de precizie 250 ohm - convertește semnalul din 4-20 mA în 1-5V |
| ACADP-20 | Sursă de alimentare AC/DC - intrare 100-240V AC, ieșire 24 V DC |
| GS-TH | Senzor de temperatură și umiditate |
| GS-LXUV | Senzor măsură intensitate luminoasă / UV, cablu 20cm |
| GS-CO2 | Senzor concentrație CO2, cablu 20 cm |
| GS-3AT | Senzor accelerație și temperatură, cablu 20 cm |
| GS-4TSR | Terminal intrare 4 x termisor, cablu 20 cm |
| GS-103AT-4P | Senzor temperatura cu termistor, tip normal, (-40 ~ 105 ºC), 3m, 4 buc/set |
| GS-103JT-4P | Senzor temperatura cu termistor, tip ultra-subțire, (-40 ~ 120 ºC), 3m, 4 buc/set |
| GS-DPA-AC | Adaptor montaj senzor curent, cablu 20 cm |
| GS-AC50A | Senzor curent (clampmetru) 50A, cablu 20 cm |
| GS-AC100A | Senzor curent (clampmetru) 100A, cablu 20 cm |
| GS-AC200A | Senzor curent (clampmetru) 200A, cablu 20 cm |
| GS-4VT | Adaptor intrare 4 canale temperatură(termocupluri) sau tensiune, cablu 20 cm |
| GS-EXC | Cablu de extensie senzor/adaptor, lungime 1.5m |
| GS-DPA | Adaptor 2 porturi pentru conectare 2 module senzor |
| B-570 | Adaptor montare GL-840 pe șină DIN |
| B-540 | Adaptor montare bază pentru terminal (B566) pe șină DIN |
| B-513 | Cablu intrare/ieșire pentru seria GL, 2m lungime |
| B-514 | Cablu alimetare DC, 2m lungime |
| SC-MAV8 | Bloc cu 8 canale pentru conversie pasivă 4-20 mA / 1-5 V |
| Documents | |
| Datasheet (EN) | Brosura si specificatii |
Industrial software
We provide industrial data acquisition and control software, custom made for customer's needs
MoreDesign and engineering
Our company provides turn-key solutions, including design of complex industrial equipment ...
MoreATEX authorization
Our company is certified by INSEMEX for engineering, installation, maintenance and inspection of Ex area installations ...
View certificate